Work with hazardous biological agents requires the highest safety precautions. Laboratories at biosafety levels BSL-2 and BSL-3 must meet strict structural and technical requirements to protect staff and the environment. For a long time this was regarded as the domain of involved solid construction. Yet modern container laboratories prove: the highest safety is also possible in modular construction.
Fundamentals of biological biosafety levels
Biological biosafety levels — known internationally as Biosafety Levels (BSL) — classify laboratories according to the hazard of the organisms handled there. In Germany these are governed by the Biological Agents Ordinance (BioStoffV) and the Technical Rules for Biological Agents (TRBA).
Low risk
Organisms with no known hazard to healthy adults
Moderate risk
Organisms with moderate hazard potential
High risk
Organisms that can cause serious disease
Highest risk
Highly hazardous pathogens with no treatment option
While BSL-1 and BSL-2 laboratories are comparatively common, BSL-3 laboratories present a particular challenge because of their complex structural requirements. BSL-4 laboratories — the highest biosafety level for work with pathogens such as Ebola or Marburg virus — have not to date been realised as a container module because of their extreme requirements.
BSL-2 in container format: proven and reliable
Laboratories at biosafety level 2 are intended for work with organisms that can cause disease in healthy adults, but for which effective treatment options exist. Typical applications are routine microbiological testing, work with cell cultures or diagnostic laboratories.
The structural requirements for BSL-2 laboratories include, among others:
BSL-2 requirements
- Enclosed laboratory area with access control
- Hand-wash basin in the laboratory
- Surfaces that are easy to clean and disinfect
- Windows must be able to remain closed during work
- Suitable biosafety cabinets for aerosol-generating activities
- Decontamination facilities (autoclave in the building)
- Separate protective clothing for the laboratory area
These requirements can be met without difficulty in a well-planned container module. Our BSL-2 containers are equipped with washable wall surfaces, integrated hand-wash basins and connections for biosafety cabinets. Technical planning takes all details into account from the start.
Expert tip: plan in a pass-through autoclave
For efficient workflows we recommend integrating a pass-through autoclave. Contaminated materials can then be decontaminated without having to be transported through the clean area. Our laboratory equipment experts advise you on the optimal configuration.
BSL-3: the premier class in container construction
Laboratories at biosafety level 3 serve work with organisms that can cause serious disease and for which effective therapies or prophylaxes are not always available. These include, for example, Mycobacterium tuberculosis, West Nile virus or certain influenza strains. The structural requirements are correspondingly strict.
Access control
Multiply secured airlocks with interlock function
Negative-pressure holding
Permanent negative pressure prevents air escape
Gas-tight cabin
Fully sealed room envelope
HEPA filtration
Extract filtration with H14 HEPA filters
The negative-pressure zone: heart of safety
The most important safety feature of a BSL-3 laboratory is controlled negative-pressure holding. Air pressure in the laboratory area sits about 30 to 50 pascals below ambient pressure. As a result, air flows inwards from outside at every opening — never the reverse. Potentially contaminated air cannot therefore reach the surroundings.
In a container laboratory this negative-pressure zone is realised by a gas-tight cabin construction. All penetrations — for electrics, media, ventilation — are executed with specialist sealing systems. The ventilation plant is designed redundantly and fitted with automatic emergency shutdown in the event of faults.
Important: observe the permit procedure
Operation of a BSL-3 laboratory requires a permit under the Biological Agents Ordinance. The approval procedure can take several months. Start preparing early — our consulting experts support you in this.
The airlock concept
Access to the BSL-3 area is via a multi-stage airlock system. Typically this consists of a black-and-white separation with changing area, a shower passage and the actual personnel airlock. The airlocks are fitted with interlocks so that both doors can never be opened at the same time.
In a compact container laboratory this airlock concept must be implemented cleverly. Optimised layouts make it possible to accommodate all required functions in a tight space. Depending on requirements, the airlocks can be housed in a separate upstream container or in the same module.
Technical systems for maximum safety
Ventilation and filter systems
The ventilation plant of a BSL-3 container laboratory is built redundantly. Two independent extract fans ensure continuous operation even if one system fails. Extract air is passed through HEPA filters of class H14, which retain at least 99.995 per cent of all particles with a size of 0.1 to 0.3 micrometres.
Supply air is also filtered to prevent the introduction of contamination. Airflow is designed so that clean air enters in the entrance area and is extracted towards the contamination source — a directed airflow that minimises the spread of aerosols.
Expert tip: bag-in/bag-out filter housings
For safe filter changes we recommend bag-in/bag-out (BIBO) filter housings. These enable contamination-free exchange of HEPA filters without exposing service personnel. Our laboratory equipment includes all required safety systems.
Decontamination and waste disposal
All materials leaving the BSL-3 area must be decontaminated. Pass-through autoclaves are typically used for this, forming the transition between the contaminated and clean areas. Liquid waste is inactivated chemically or thermally before discharge.
For room gassing — for example after an incident or before maintenance work — connections for formaldehyde evaporators or hydrogen peroxide generators must be provided. These enable complete decontamination of the entire laboratory area.
Monitoring and documentation
A BSL-3 laboratory requires extensive monitoring and documentation systems. Pressure sensors continuously monitor negative-pressure holding and trigger an alarm if there are deviations. Door contacts log every access. Video surveillance documents activities in the laboratory.
Our Smart Lab solution for BSL-3 includes:
- Continuous differential-pressure measurement with alarm system
- Automatic documentation of all safety-relevant parameters
- Access control with biometric capture
- Video surveillance with recording
- Remote access to status data for authorised persons
- Automatic notification to the operations manager in the event of faults
Advantages of the container concept for high-containment laboratories
Realising BSL-2 and BSL-3 laboratories in container form offers specific advantages that go beyond the general merits of modular construction:
- Controlled manufacture: The gas-tight cabin is created under workshop conditions with the highest precision — quality levels that would barely be achievable on a construction site.
- Independence from the existing building: A container BSL-3 can be installed without involved interventions in existing building structures.
- Rapid availability: Realisation time is significantly shorter than for conventional BSL-3 laboratories, which often take years.
- Flexibility: If the site changes, the complete laboratory can be moved — the entire certified unit remains intact.
- Defined interfaces: Clear separation between high-containment area and surroundings eases operation and maintenance.
From concept to approval
Realisation of a BSL-2 or BSL-3 container laboratory follows a structured process. At the beginning sits the needs analysis: which organisms are to be processed? Which working procedures are planned? How many people will work in the laboratory? On the basis of this information we develop the concept.
In parallel, coordination with the permitting authorities begins. For BSL-2 laboratories a notification to the competent authority is generally sufficient. BSL-3 laboratories require a permit, the granting of which depends on a review of the structural and organisational prerequisites. Our consulting experts accompany you through this process.
Conclusion: safety and flexibility united
The question “BSL-2 and BSL-3 in a container — is it possible?” we can answer with a clear yes. Modern manufacturing technologies and well-thought-out concepts make it possible to realise high-containment laboratories in mobile construction that stand comparison with conventional facilities in every respect. On the contrary: controlled workshop manufacture offers quality advantages, while the flexibility of the container concept opens new fields of use.
Enquire about a containment laboratoryAbout the author